ORALIX AC SYSTEM
PREVENTIVE MAINTENANCE
4.2.4. Maintenance of the Wall Support
For the maintenance of the Wall Support follow the
procedure below:
Remove the cover (c) and check that the Wall
Support is firmly connected to the wall by four
fixation points (the applied torque shall be 10 N/m).
Check also that it is properly levelled.
Check the power connections at the terminal block
(d) and inspect the protective earth connection (e)
at the ground point, checking for damage or corrosion
which could alter the electrical continuity.
(c)
(d)
(e)