4519-190-70943_4
D-1
ORALIX AC SYSTEM
ACCEPTANCE
Gendex Dental Systems
SECTION D
ACCEPTANCE
1.
INTRODUCTION ............................................................................................... D-1
2.
EQUIPMENT REQUIRED ................................................................................ D-1
3.
ACCEPTANCE TESTS .................................................................................... D-2
3.1.
LABELS, INDICATORS AND WARNINGS ..................................................................................... D-2
3.2.
ACCURACY OF kVpeak ............................................................................................................... D-2
3.3.
ACCURACY OF TUBE CURRENT ................................................................................................ D-2
3.4.
ACCURACY OF EXPOSURE TIME ............................................................................................... D-2
3.5.
RADIATION LEAKAGE ................................................................................................................. D-3
3.6.
EARTH RESISTANCE ................................................................................................................... D-3
3.6.1.
Wall mounted systems ................................................................................................... D-3
3.6.2.
Mobile systems .............................................................................................................. D-3
3.7.
EARTH LEAKAGE ........................................................................................................................ D-4
DRAWINGS:
ACCEPTANCE CHECKLIST ................................................................................................................. DZ-1
1. INTRODUCTION
This section contains information about acceptance tests, required to ensure the customer that the Oralix AC
system conforms to the specification.
2. EQUIPMENT REQUIRED
Digital Multimeter (Accuracy
=/<
1%).
Oscilloscope.
Bonding Tester.
No-invasive KV meter (Accuracy
=/<
2%).