Probes
13-30
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
The instructions provided above have been validated to properly
prepare GE Ultrasound probes for re-use. It remains the
responsibility of the processor to ensure that the processing, is
performed as specified in this document. This may require
verification and routine monitoring of the process.
Chemicals Used for Efficacy Validation
The table below lists the products and intended use (clean, ILD,
HLD) that were validated.
Table 13-4: Chemicals used for Efficacy Validation on non-TEE probes
Product
Type
Trade Name
Manufacturer
Minimum
Contact
Time
Active
Ingredient
Validated for
probe
Enzymatic
Detergent
(Soak)
Enzol®
(Cidezyme®)
Advanced
Sterilization
Products®
(J&J)
1-Minute
Soak
Proteolytic
Enzymes
3Sc-RS,
M5Sc-D, 6S-D,
12S-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
MetriZyme™
Metrex™
Prolystica® 2X
Concentrate
Presoak &
Cleaner
Steris
Intermediate-l
evel
Disinfectant
(wipe)
Oxivir® Tb
Diversey
10-Minute
Exposure
Hydrogen
Peroxide
3Sc-RS,
M5Sc-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
High-level
Disinfectant
(Soak)
Cidex® OPA
Advanced
Sterilization
Products (J&J)
10-Minute
Soak
Ortho-
phthalaldehyde
3Sc-RS,
M5Sc-D, 6S-D,
12S-D, 9L-D,
11L-D, C1-5-D,
C1-6-D, C2-9-D,
C3-10-D,
iC5-9-D,
L8-18i-D
McKessen
OPA/28
McKesson
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......