Measurements and Analysis
8-98
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Doppler alignment
Errors in velocity measurements increase with the cosine of the
angle between the measured flow and the ultrasound beam. For
example, an alignment error of 20 degrees, will give a 6%
under-estimation of the velocities, while an error of 40 degrees
will cause the under-estimation to be 24%. It is highly
recommended to optimize transducer position to align the
ultrasound beam with the flow direction.
NOTE:
If alignment is not possible, you may use the Angle Correction
control to compensate if the flow direction is known.
Screen pixel resolution
The display screen is composed of an array of square picture
elements (pixels). The smallest resolvable unit is +/- 1pixel. This
pixel error is only significant when measuring short distances on
the screen. By observing good scanning practices, the settings
of the field of view should be such that the measured distance is
significant with respect to the full size of the screen. When such
scaling is impossible, the pixel error may come into play. The
pixel error is +/- 0.2% of the full ultrasound area in the User
Screen.
Algorithms
Some formulae used in clinical calculations are based on
assumptions or approximations. For example the volume
calculations from 2D or M mode assume a certain, ‘ideal’ shape
of the heart chamber, while the actual shape can vary quite
much between individuals. Also, formulae taking several “raw”
measurements as inputs are prone to increased errors,
depending on the combination of input variable accuracies. For
example, the Cardiac Output formula from Doppler is sensitive
to errors in the entered Diameter, since this parameter will be
squared in the formula.
Speed of Sound in Tissue
The average speed of sound value of 1540 meters / second is
used for all calculations. Depending on the tissue structures, this
generalization may give errors from 2% (typical) to 5% (much
fatty tissue layers present).
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......