Image Optimization
5-50
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Frequency
2D, M-Mode
Rotate to adjust the probe's operating frequency. The selected frequency is
displayed in the
Parameter
window. For some probes/applications the lowest
frequency settings will be Octave imaging settings.
Push to turn Octave imaging on/off.
CF, Doppler, TVI,
SRI
Enables the adjustment of the transmission frequency to control the sensitivity or
the level of penetration. The selected frequency is displayed in the
Parameter
window. Adjusting Frequency may affect Sample Volume and LVR settings.
Frame rate
2D, CF, Doppler, TVI,
TT, SR, SRI, TSI
Adjusts frame rate (FPS). The relative setting of the frame rate is displayed in the
Parameter
window. When adjusting frame rate, there is a trade off between spatial
and temporal resolution.
Up/Down
2D, M-Mode, CF
Up/Down
: enables the 2D image to be flipped 180 degrees.
Left/Right
2D, CF
Left/Right
: enables a mirror image of the 2D image to be created. The left/right
reference marker
V
moves to the other side of the image.
Compress
2D, M-Mode
Controls the amount of contrast in the 2D image. An index number is displayed on
the control to indicate the relative level of compression.
Doppler
Enables control over the contrast of the Doppler spectrum. When compression is
raised, the spectrum image becomes softer and some low level background noise
may appear. Compress is available in both Live and Freeze.
TVI, SRI
Controls the amount of color compression. The color bar is adjusted accordingly.
Reject
2D, M-Mode
Adjusts the rejection level. When this control is increased, low-level echoes are
rejected and appear darker in the 2D image. An index number is displayed on the
control to indicate the relative level of rejection.
Doppler
Enables undesirable background noise to be removed from the Doppler spectrum
resulting in a darker background. Reject is available in both Live and Freeze.
Data Dependent Processing (DDP)
2D
Performs a temporal processing which reduces random noise without affecting the
motion of significant tissue structures. An index number is displayed on the control
to indicate the relative DDP level.
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......