Image acquisition
Vivid S70 / S60
–
User Manual
7-11
BC092760-1EN
01
Running Continuous capture
1. Do all your pre-stress acquisitions in the Cardiac
application.
2. Press
Stress
on the Touch panel to enter the stress echo
mode.
The
Protocol
screen is displayed (Figure 7-1
3. Press
Template
.
The template list is displayed.
4. Select the template
Exercise 2x4
.
5. Press
Begin/Cont
.
6. Acquire the resting loops in all four views.
7. Once the fourth loop is acquired the system enter into a
waiting mode where Continuous Capture is in pause state
awaiting the patient to exercise.
8. When the patient is back on the bed, press
Store
or
2D
Freeze
. The Continuous capture acquisition is started.
9. Acquire all your views. You may pause the acquisition when
moving the probe between parasternal and apical position.
The memory buffer gauge increases (Figure 7-4). When
memory filling exceeds 90%, the percent number turns red.
10. Press
Freeze
to finish.
11. Press
Select Cycles
.
The
Continuous capture selection
screen is displayed (see
).
If the buffer is filled up the system will automatically display
the
Continuous capture selection
screen.
Refer to the next section if additional image acquisition is
necessary after the buffer is filled up.
12. Assign the cineloops to the four views (see page 7-13).
The continuous capture acquisition was stored on disk
before the
Continuous capture selection
screen was
displayed. At end exam, a dialogue is displayed asking
whether this file should be kept or not. To avoid
unnecessary data storage, it is recommended to delete this
file at end exam when the cycles have been selected and
stored.
13. Press
Done
if you are finished selecting or want to do the
selection later.
OR
Press
Select later
if you want to return to the scanning
screen. If there are more rows in the protocol template, the
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......