Measurement accuracy
Vivid S70 / S60
–
User Manual
8-101
BC092760-1EN
01
dt
0.025 - 1.0 s
0.5%
Event timing
Time
50 - 840 ms
max (5.0%, 1 ms)
Q Analysis
Velocity from TVI data
12 - 29 cm/s
20%
Strain rate from TVI data
-3 - +4 s
-1
0.6 s
-1
Strain from TVI data
-33 - +51%
6.6 percentage points
Displacement from TVI data
-0.9 - +1.7 mm
0.24 cm
2D Auto EF
EF
30 – 80%
15 percentage points
ESV
20 – 120 ml
35 ml
EDV
20 – 120 ml
45 ml
Ls and Ld
8 – 10 cm
2 cm
EF_BiP
30 –80%
15 percentage points
ESV_BiP
20 – 120 ml
20 ml
EDV_BiP
20 – 120 ml
20 ml
AFI
Global Longitudinal Strain
-5 – -25%
3 percentage points
Regional Longitudinal Strain
0 – -25%
9 percentage points
2D Strain endocardial/
epicardial strain
-5 – -25%
9 percentage points
4D Auto LVQ
Volumes (EDV, ESV)
10 – 300 ml
10 ml
The tool should not be used
on collapsing ventricles
Ejection Fraction
15 – 85%
15 percentage points
Sphericity Index
0.1 – 1.5
0.2
4D Auto MVQ
Annulus Area 2D
4.1 - 22.8 cm
2
10%
Accuracy w/o editing: 15%
Annulus Area 3D
4.6 - 25.3 cm
2
10%
Accuracy w/o editing: 15%
Annulus Height /
MV Annulus Height
1.7 - 10.5 mm
max (10%, 3 mm)
Accuracy w/o editing: 5 mm
Measurement
Range
Accuracy
Comments
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......