Measurements and Analysis
8-54
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
End-diastolic volume contour detection
1. Select
EDV
(End-Diastolic Volume).
The default screen displays three apical views, three short
axis views and an interactive view (Figure 8-28).
The system automatically displays the loops at the
estimated end diastolic frame.
2. If the automatic end diastolic frame detection is not optimal,
use the
Move ED
control to set the new end diastolic frame.
3. In one of the apical views, place two points: one at the
center of the LV base and one at the apex.
A contour is automatically drawn in all views (see
Figure 8-28).
NOTE:
Alternative methods for doing contour detection: press
Auto
on the Control panel to draw the contour automatically
without placing any points, or press
Manual
on the control
panel and place two basal and one apical point in all three
apical views.
NOTE:
Erroneous contour detection of the left ventricle may lead to
incorrect measurement results. The contour detection
should be visually checked and edited if required.
The contour detection should be checked in all slices.
NOTE: Press
Layout
several times on the control panel to display
the apical slices in large size one by one.
The contour detection quality can be visualized by moving
the cursor through the LV and observe the contour detection
in the interactive view. The position of the interactive view
updates according to the cursor position in the other views.
To lock the interactive view, press
Lock view
. When the
interactive view is locked the position can be controlled
using the
Move view
control.
Adjust
Contour visibility
to change the display intensity of
the contour (or hide it).
To run the cineloop, press
2D Freeze
on the Control panel
or the rotary button
Run
on the Touch panel.
To stop the cineloop, press
2D Freeze
on the Control panel
or the rotary button
Go To ED
on the Touch panel. The end
diastolic frame is displayed.
To edit the contours, see ‘Contour adjustment’ on
.
To reset the contours and start over, press
Reset
.
The following measurements are available:
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......