Color Mode
Vivid S70 / S60
–
User Manual
5-17
BC092760-1EN
01
NOTE:
The scale value may affect FPS, Low Velocity Reject, and
Sample Volume.
•
Adjust
Low Velocity Reject
to remove low velocity blood
flow and tissue movement that reduces image quality.
•
Adjust
Variance
to detect flow disturbances.
•
Adjust
Sample volume
(SV) to a low setting for better flow
resolution, or a higher setting to more easily locate disturbed
flows.
•
Adjust
Frequency
to optimize the color flow display. Higher
settings improve resolution. Lower settings improve depth
penetration and sensitivity. This does not affect the
frequency used for 2D and M-Mode.
NOTE:
Frequency setting may affect FPS, SV and Low Velocity
Reject.
In certain selected applications when entering color flow
imaging (CF), there is a button available called
AutoFrequency
. If this button is depressed, the application
will automatically select for the user a frequency appropriate
for the depth of the ROI. When the feature is enabled, the
frequency control is no longer available to the user until the
AutoFrequency button is released again. And while the
AutoFrequency feature is active the scale control is
controlling velocity and not PRF as indicated on the touch
panel rotary itself. If the user turns the desired velocity
control to a higher value than what is achievable at the
maximum PRF attainable at the current depth, the feature
will automatically lower the frequency to still attempt to
achieve the requested velocity, if this is possible within the
range of frequencies supported. Moving the ROI between
different depths, the user may experience that the system
selects a different frequency if that is more suitable, as is
displayed on the screen, but will attempt where possible to
retain the velocity span requested from the user interface.
Releasing the AutoFrequency button will again make the
mnual frequency selection rotary available to the user and
the scale button will again be controlling the PRF directly.
•
Adjust
Power
to obtain an acceptable image using the
lowest setting possible.
NOTE:
The Power setting affects all other operating modes.
Содержание Vivid S70
Страница 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Страница 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Страница 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 140: ...Image Optimization 5 4 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 2 2D Touch panel 4D probe Live page 1 and 2...
Страница 144: ...Image Optimization 5 8 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 4 M Mode Touch panel page 1 and 2...
Страница 149: ...Color Mode Vivid S70 S60 User Manual 5 13 BC092760 1EN 01 Figure 5 6 Color 2D Touch panel page 1 and 2...
Страница 156: ...Image Optimization 5 20 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 10 The PW Doppler Touch panels page 1and 2...
Страница 160: ...Image Optimization 5 24 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 12 TVI Touch panel page 1 and 2...
Страница 163: ...Tissue Tracking Vivid S70 S60 User Manual 5 27 BC092760 1EN 01 Figure 5 14 The Tissue Tracking Touch panel page 1 and 2...
Страница 166: ...Image Optimization 5 30 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 16 The Strain rate Touch panel page 1 and 2...
Страница 169: ...Strain Vivid S70 S60 User Manual 5 33 BC092760 1EN 01 Figure 5 18 The Strain Touch panel page 1 and 2...
Страница 172: ...Image Optimization 5 36 Vivid S70 S60 User Manual BC092760 1EN 01 Figure 5 20 The TSI Touch panel page 1 and 2...
Страница 215: ...4D Mode Vivid S70 S60 User Manual 6 15 BC092760 1EN 01 Figure 6 9 The 4D Color Flow mode Touch panel...
Страница 280: ...Measurements and Analysis 8 2 Vivid S70 S60 User Manual BC092760 1EN 01 Measurement accuracy on page 8 97...
Страница 297: ...Advanced cardiac measurements and analysis Vivid S70 S60 User Manual 8 19 BC092760 1EN 01 Figure 8 9 AFI workflow...
Страница 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Страница 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 457: ...Disk management Vivid S70 S60 User Manual 10 55 BC092760 1EN 01 8 Press OK See page 10 56 to perform a database backup...
Страница 479: ...Working with the report function Vivid S70 S60 User Manual 11 5 BC092760 1EN 01 Figure 11 1 The Report screen...
Страница 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Страница 709: ......