Getting started
Vivid S5/Vivid S6 User Manual
57
R2424458-100 Rev. 2
To help assure grounding reliability, connect to a "hospital
grade" or "hospital only" grounded power outlet.
Figure 2-5: Example Plug and Outlet Configurations
Battery
The System contains an internal battery (optional), which is not
user-accessible. The internal battery acts as a limited
alternative power source, which becomes active when the AC
power fails or AC cable is abruptly pulled out.
Display turns off and no scanning is possible while in battery
mode. For further information about use of the battery see
section "Battery Power Mode" on page 63.
The battery may
only be disassem-
bled and re-assem-
bled by an
authorized
field-service engi-
neer.
Battery technology
The lithium ion technology used in the system’s battery is
significantly less hazardous to the environment than the lithium
metal technology used in some other batteries (such as watch
batteries). Used batteries should not be placed with common
household waste products. Contact local authorities for the
location of a chemical waste collection program nearest you.
CAUTION
Use caution to ensure that the power cable does not disconnect
during system use. If the system is accidentally unplugged, data
may be lost.
WARNING
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than specified.
Never use an extension cord or adapter plug.
1. 100-120 VAC, 1200 VA Plug and Outlet Configuration
2. 220-240 VAC, 1200 VA Plug and Outlet Configuration
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......