Measurement and Analysis
Vivid S5/Vivid S6 User Manual
307
R2424458-100 Rev. 2
The user may modify the offset or length while performing the
exam, in case these are not set to optimal dimensions.
To modify Protocol defaults
1.
Set the Protocol button ON or OFF.
2.
In case Protocol is ON: adjust the
Offset
and
Length
values.
3.
Press the
Store default
soft-key to store the status of
"protocol" button ON or OFF and the current settings of
Offset
and
Length
, to be used in the future exams.
4.
The default settings may be defined differently for the CCA
and the ICA.
Note
: When adjusting the offset for the ICA, the offset
values are negative, as the ROI is located to the left of the
vertical reference marker.
IMT trace approval
Since the IMT measurements are done semiautomatically, the
operator has to approve the detection by visual inspection
before storing the results in worksheet and report.
•
If the traces fit both layers of the intima-media walls,
approve the measurement by selecting
Transfer
in the
Measurement menu
.
Once transferred, the calculations can be viewed in the
worksheet and report.
CAUTION
Measurements that are not approved will not be saved.
CAUTION
Any image adjustments (e.g. Gain or zoom) on approved
(transferred) measurements will unassign the measurements.
Press
Transfer
to approve the measurements again.
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......