Measurement and Analysis
Vivid S5/Vivid S6 User Manual
275
R2424458-100 Rev. 2
Acquisition
1.
Create an exam, connect the ECG device and make sure
to obtain a stable ECG trace.
2.
Acquire 2D grey scale cineloops of an Apical long axis
(APLAX) view, an Apical 4 chamber view and an Apical 2
chamber view.
3.
Store the loop (must store).
Note
: It is recommended to acquire all three apical views
sequentially in order to get similar heart rate in all views.
•
The frame rate should be between 37 and 80 frames per
second. A high frame rate is recommended for high heart
rate.
•
The scanner should be configured to store at least 100 ms
before and after each heart cycle,
- or -
•
The scanner should be configured to store 3 loops or more.
•
If the stored loop contains more than one heart cycle, the
analysis will be done on the second last heart cycle.
•
The entire myocardium should be visible.
•
A depth range that includes the entire left ventricle should
be used.
Starting AFI
1.
Open an APLAX view and press
MEASURE
.
2.
In the
Measurement menu
, select
AFI
. The View selection
menu is displayed (Figure 7-11).
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......