Probes
544
Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Maximum probe temperature
Notes:
Lens temperature measured under following conditions
per IEC 60601-2-37 Amd. 2:
1.
Thermocouple was placed at the geometric center of the
lens. Thermal phantom made with tissue-mimicking
material as referenced in IEC60601-2-37 Amd 2: Annex
BB.
2.
a: Thermal phantom at 23 +/- 3 °C for all probes.
b: Temperature rise is measured and added to 33 for
External-Use probes or 37 °C for Non External-Use
probes.
3.
Probe placed upright in contact with above thermal
phantom.
Probe
Max Temp
3S -RS
40.5
M4S-RS
42.8
5S-RS
40.8
6S-RS
39.5
7S-RS
37.7
10S-RS
36.1
8L-RS
41.0
9L-RS
40.4
12L-RS
36.6
i12L-RS
40.5
4C-RS
37.8
8C-RS
37.3
E8C-RS
40.7
6T/6T-RS
40.8
6Tc-RS
42.3
9T/9T-RS
42.1
P2D
35.3
P6D
36.0
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......