Probes
Vivid S5/Vivid S6 User Manual
551
R2424458-100 Rev. 2
Disconnecting the probe
Probes can be disconnected at any time. It is recommended
that the probe should not be active when being disconnected.
To disconnect a probe (small connector -
type RS)
1.
Freeze the image by pressing
FREEZE
.
2.
Press the connector locking lever towards the left to unlock
the connector.
3.
Pull the probe and connector straight out of the probe port.
4.
Carefully slide the probe and connector away form the
probe port and around the right side of the keyboard.
5.
Ensure that the cable is free.
6.
Verify that the probe head is clean before placing the probe
in its storage case.
To disconnect a probe (large connector -
type OR)
1.
Rotate the lock handle counter-clockwise to the horizontal
position to unlock the connector.
2.
Pull the connector straight out of the probe port.
3.
Verify that the probe head is clean before placing the probe
in its storage case.
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......