Measurement and Analysis
Vivid S5/Vivid S6 User Manual
311
R2424458-100 Rev. 2
2.
The following controls may be adjusted from the control
panel:
•
CYCLE SELECT
: change the selected cycle.
•
TRACE SENSITIVITY
: optimize the trace contour.
•
CURSOR SELECT
: select Peak systolic or End diastolic
marker. The selected marker can be moved to a new
location. Press
SET
to anchor the marker to its new
location.
Assigning auto calculations
1.
In the
Vascular Measurement menu
(Figure 7-37), select:
• The actual vessel name
•
Prox
,
Mid
or
Dist
: the location of the vessel (Proximal,
Middle or Distal).
•
Rt
or
Lt
: right or left side of the patient.
2.
Press
Transfer
to assign the measurements and
calculations.
To undo the assign-
ment, press
Cancel
transfer
.
The
Measurement result table
is updated accordingly and
the measurements and calculations are added to the
worksheet and report.
Manual vascular calculation
When doing manual measurements, the system can detect the
trace automatically or it can be drawn by the user. This is
controlled by the
Auto
and
Manual
commands is the
Measurement menu
.
The following instructions assume that you first scan the patient
and press
FREEZE
.
1.
Adjust the vessel location parameters in the
Vascular
measurement menu
(Figure 7-37).
2.
Select the measurement to be performed from the
Measurement menu
or from the
Show All menu
for
additional measurements.
3.
Perform the measurement as described below.
Acceleration, Acceleration time (AT)
1.
Select
Accel
or
AT
.
2.
Position the caliper at the start point and press
SET
to
anchor the caliper.
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......