Measurement and Analysis
Vivid S5/Vivid S6 User Manual
329
R2424458-100 Rev. 2
• To fix the start point, press
SET
.
The system fixes the first caliper and displays a second
active caliper.
• To position the second active caliper at the end point,
move the Trackball.
A dotted line connects the measurement points.
• To complete the measurement, press
SET
.
The system displays the distance value in the
Results
Window
.
3.
When the measurement of the first quadrant is completed,
unfreeze and move to the second quadrant.
4.
After you obtain the image, press
FREEZE
and then
MEASURE
.
The system prompts you to continue with the
AFI
measurements. Make sure that the next quadrant has
been selected.
5.
Perform a standard distance measurement for the second,
third, and fourth quadrants (see step 2 above).
6.
When all four quadrants have been measured, the system
calculates the AFI total and displays it in the
Results
Window
.
Note
: If the fluid in a pocket is zero, set the second caliper on
top of the first one to give it a zero value.
Figure 7-44: Amniotic Fluid Index
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......