Safety
Vivid S5/Vivid S6 User Manual
45
R2424458-100 Rev. 2
electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the
measured field strength in the location in which the ultrasound
system is used exceeds the applicable RF compliance level as
stated in the immunity declaration, the ultrasound system
should be observed to verify normal operation. If abnormal
operation is observed, additional measures may be necessary,
such as re-orienting or relocating the ultrasound system or
using an RF shielded examination room may be necessary.
1.
Use either power supply cords provided by GE Medical
Systems or ones designated by GE Medical Systems.
Products equipped with a power source plug should be
plugged into the fixed power socket which has the
protective grounding conductor. Never use any adaptor or
converter to connect with a power source plug (e.g.
three-prong-to-two-prong converter).
2.
Locate the equipment as far away as possible from other
electronic equipment.
3.
Be sure to use only the cables provided by or designated by
GE Medical Systems. Connect these cables following the
installation procedures (e.g. wire power cables separately
from signal cables).
4.
Lay out the main equipment and other peripherals following
the installation procedures described in the Option
Installation manuals.
General notice
1.
Designation of Peripheral Equipment Connectable to This
Product.
The equipment indicated in Chapter 12, "Peripherals" on
page 575 can be hooked up to the product without
compromising its EMC performance. Avoid using
equipment not designated in the list. Failure to comply with
this instruction may result in poor EMC performance of the
product.
2.
Notice against User Modification The user should never
modify this product. User modifications may cause
degradation in EMC performance. Modification of the
product includes changes in:
• Cables (length, material, wiring, etc.)
• System
installation/layout
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......