Archiving
444
Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Scanner’s connectivity configuration
The scanner’s TCP/IP settings must be configured as
described in "Scanner’s TCP/IP settings" on page 438.
In addition, to work against the DICOM server the following
information has to be entered in the scanner:
•
The DICOM server IP address
•
The DICOM server port number
•
The DICOM server AE title (the server’s name)
Typically source for this information in the network
administrator.
Setup of the DICOM server in the scanner’s configuration
management package
DICOM dataflow selection
1.
Press
CONFIG
and log on as administrator.
2.
Select the
Connectivity
category and
Dataflow
subgroup.
The
Dataflow sheet
is displayed.
3.
Select the DICOM dataflow to configure in the
Name
pull-down menu
(see Figure 9-29).
Figure 9-29: The Dataflow sheet
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......