Stress Echo
202
Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Figure 5-3: Display of the Reference loop during acquisition
Protocol Pause function
During the stress acquisition it is possible to temporarily exit the
protocol acquisition mode to acquire images in any mode
outside the stress protocol.
1.
To temporarily exit the protocol mode, press
PROTO.IN/OUT
soft-key button.
2.
Acquire the desired images outside the protocol.
3.
Press the
2D
mode button.
4.
Press
PROTO.IN/OUT
soft-key button to restart the protocol
acquisition mode and resume the stress acquisition.
Selecting a view during acquisition
A fixed protocol is provided for scanning, based on the selected
template. The system automatically highlights the next view to
be acquired in the template matrix, as images are stored.
1. Current acquisition loop
2. Corresponding reference loop
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......