Measurement and Analysis
294
Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Figure 7-25: Measurement and View selection menus
3.
Select the name of the current view:
4-ch
or
2-ch
. A ROI
can be defined.
4.
Place 3 points, 2 on the basal area and one on the apex
according to the guidance.
Tracing the endocardial border
When selecting a view to analyze, the system automatically
displays a frame where the endocardial border is usually
clearly visible. To use another frame, adjust
SELECT FRAME
.
To trace the endocardial border, place three points at the
endocardial border; two annular points at the base and one at
the apex. Follow the indications displayed on the screen when
placing the three points.
After placing the third point on the Apex an endocardial border
will automatically be traced (Figure 7-26).
Note
: Correct border tracing is important for an accurate EF
measurement. The system has an adaptive border tracing
function: using the endocardial three points as a guide, the
system will analyze the image and automatically adapt the
border tracing to an optimal position.
Note
: the Yo-yo function is turned on to help finding correct
location for the points.
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......