Scanning Modes
190
Vivid S5/Vivid S6 User Manual
R2424458-100 Rev. 2
Alternative assignable controls
Press
ALT
to access to the following modes:
•
Curved Anatomical M-Mode (see page 155)
• Tissue
Synchronization Imaging (see page 188)
•
Tissue Tracking (see page 173)
•
Strain rate (see page 178)
•
Strain (see page 183)
Physio assignable control
Press
PHYSIO
and
MORE
twice to access to the following
control:
QRS visible
Shows/hides the QRS marker on the ECG.
TSI Soft menu controls
TSI Cut-off
Controls the cut-off time: using this control it is possible to color
all parts of the TSI image that has a time to peak less than a
certain cutoff time.
Threshold
Controls the level of greyscale intensity that is used as
threshold for color.
Transparency
Control the degree of transparency of the TSI color.
Frequency
Enables the adjustment of the transmission frequency to
control the sensitivity or the level of penetration.
Power
Controls the amount of acoustic power applied. When power is
reduced the signal to noise ratio is also reduced, so that the
image may become noisier.
Cine Compound (Freeze only)
Calculates and displays cineloops generated from a temporal
averaging of multiple consecutive heart cycles. The number of
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......