Measurement and Analysis
Vivid S5/Vivid S6 User Manual
259
R2424458-100 Rev. 2
M-Mode Measurements
In M-Mode, the user can perform distance and time
measurements. This measurement package has also the
following pre-defined measurement studies:
• LA/Ao
• LV
• RV
M-Mode Length measurements
1.
Generate the M-Mode image.
2.
Press
FREEZE
to stop the cineloop.
Alternative: Press
CALIPER
on the con-
trol panel and press
CALIPER
assignable.
3.
Press
MEASURE
on the Control Panel.
4.
Select
caliper
in the
Measurement Menu
.
5.
Trackball the cursor to the start point of the measurement.
6.
Press
SET
to anchor the start point of the measurement.
See the
Status bar
to get the next step
to perform.
7.
Trackball the cursor to the measurement end point.
The current distance value is displayed in the
Measurement result table
and is instantaneously updated
when moving the cursor.
8.
Press
SET
to anchor the end point of the measurement.
The measurement
display color on the
M-Mode changes
from green to blue
after completion of
the measurement.
The measurement result is displayed in the
Measurement
result table
.
9.
To assign a label to the measurement, see page 252.
10. Repeat steps 5 through 8 to make additional length
measurements.
The measurements
displayed on the
M-Mode image and
the corresponding
results are num-
bered.
Editing M-Mode Length measurements
1.
Trackball the cursor to one of the anchor points of the
measurement to modify.
2.
Press
SET
twice (double-click).
The selected marker turns green and is unanchored.
3.
With the
Trackball
, reposition the marker to a new position.
4.
Press
SET
.
Содержание Vivid S5
Страница 18: ...Revision History xvi Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 30: ...Introduction 12 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 154: ...Basic scanning operations 136 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 250: ...Stress Echo 232 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 260: ...Contrast Imaging 242 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 420: ...Quantitative Analysis 402 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 508: ...Archiving 490 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 600: ...Peripherals 582 Vivid S5 Vivid S6 User Manual R2424458 100 Rev 2 ...
Страница 689: ......
Страница 690: ......