GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm
Page no. 191
Chapter 16
H30 Please turn off
control console
The system cannot shut down the
generator and X-ray console.
- case A. Following an operator
request on the browser.
- case B. Following a generator
error that requested generator
shutdown.
A system shut down should be
requested manually.
1) Power OFF on the X-ray console (immediately
in case B!).
2) If the error occurred during a normal shut down:
A. Note ALL messages and contact GEMS Field
Service to inform the field engineer that the system
should be shut down itself.
B. You can continue to use the system but you
must power OFF from the X-ray console at the
next occurrence of this error.
H31 Positioner power
supply failure
The system is set in emergency
mode:
- Gantry power supply failure; the
system is set in a "halt" state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H32 Tube head control
failure
The system is in emergency
mode:
- Control command from tube
head has failed, the system is set
to "halt" state.
- It is not possible to perform any
movement or image acquisition.
1) Uncompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H34 Tube head com-
munication failure
The system is in emergency
mode:
- Tube head control is not commu-
nicating, the system is set to "halt"
state.
- It is not possible to perform any
movement or image acquisition.
1) Uncompress the patient using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
W01 Internal software
error
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
W02 Internal software
error
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
W03 Internal software
error
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
W04 Internal software
error
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
W05 Internal software
error
An error occurred on the Gantry
CPU that does not affect the cur-
rent system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
Message on
Gantry Readout
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).