GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm
Page no. 169
Chapter 16
I14 AWS was not ready
AWS system was not ready to
perform an image acquisition.
1. Press the prep button again.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
I15 Console locked
The X-ray Console is locked due
to service access (detector cali-
bration or IQ tool is in progress).
Actions on right-hand section of
X-ray Console are still available.
Wait until end of service access.
If no service in progress, power OFF/ON from the
X-ray Console.
I16 Collimator locked
The diaphragm control is locked
because detector calibration is in
progress.
Wait until end of calibration.
If no calibration in progress, power OFF/ON from
the X-ray Console.
I17 AWS/SENO comm.
lost
AWS system cannot communi-
cate with Generator.
A service laptop may be con-
nected in place of the AWS, or
the AWS may be powered OFF
for service requirements.
1. Wait for completion of the service operation if
there is one, or wait one minute to allow the system
to recover.
2. If the problem persists, power OFF the whole
system and wait for the end of the shutdown
sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
I18 Collimator/Console
locked
The X-ray Console and the dia-
phragm control are locked due to
service access (detector calibra-
tion in progress). Actions on
right-hand section of X-ray Con-
sole are still available.
Wait until end of service access.
If no service in progress power OFF/ON from the
X-ray Console.
I19 IDC/SENO comm.
lost
The Image Detection Controller
system cannot communicate with
the Generator.
If communication is not recov-
ered in a few minutes, the system
will be in failure mode (error
E19).
Wait until the Image Detection Controller system
recovers.
I21 Heating failure track
1
Molybdenum focal track is out of
use.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
You can continue using rhodium focal track.
I22 Heating failure track
2
Rhodium focal track is out of use. Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
You can continue using molybdenum focal track.
Message on
X-ray Console
Explanation
Corrective Action