Chapter 8
Quality Check
Aim of the Quality Check feature
The Quality Check feature requires the user to apply a qualification level to each image. This qualifica-
tion level is:
•
Used to trigger Auto Push and/or Auto Print
•
Used to generate image quality statistics over a given period of time (Repeat and Reject Analysis or
RRA)
How to configure Quality Check
Initial decision: Do you need the RRA feature?
The RRA feature provides an alternative (and automated) method to the manual Repeat Analysis
method described in the Senographe DS QC Manual.
It is mainly intended for sites that follow MQSA regulations. For such sites, it is mandatory to perform a
repeat analysis at least quarterly.
It is also useful for other sites to generate and keep statistics on total repeat and reject rates, as well as
their respective reasons.
Case 1: RRA enabled
If you need the RRA feature, click Yes in the Enable Repeat and Reject Analysis module, then select the
cases for which you wish to make use of the Auto Print and Auto Push features.
Note:
Once RRA is enabled, the Default image quality check is set to OK and cannot be modified.
Case 2:
RRA disabled
Case 1:
RRA enabled
Save
Close
Default image quality check
OK
NOT OK
Enable Repeat and Reject Analysis
Auto Push Repeated images?
Yes
No
Auto Push Rejected images?
Auto Print Repeated images?
Auto Print Rejected images?
Auto Push
Auto Print
Annotations
Auto Windowing
Quality Check
Auto Delete
Test Hosts
Names
Image Process.
MedAppPrefs
Yes
No
Yes
No
Yes
No
Yes
No