Chapter 16
Page no. 166
16-errors.fm
GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
Table 1 Error Messages displayed on the X-ray Console
Message on
X-ray Console
Explanation
Corrective Action
E01 Generator failure
Problem reported in HV section
of generator.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E02 Focus bias failure
Problem reported in tube focus
bias circuits.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E03 Rotor failure
Problem reported in anode
starter.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E04 Arm fatal failure
Generator detects error in dialog
with arm.
Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
E05 Failure during
exposure
Current exposure aborted due to
failure in equipment.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
inform GEMS Field Service.
E06 Software error (gen
CPU)
An error occurred in the Genera-
tor CPU software.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
inform GEMS Field Service.
E07 Please TURN OFF
the console !
A generator emergency failure
occurred but a second failure
prevented the system from shut-
ting itself down.
1) Note ALL messages.
2) Switch OFF the system from the X-ray console
to quickly switch off the generator.
3) Contact GEMS Field Service.
E08 Power supply fail-
ure
Supply of generator power cir-
cuits is not authorized.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E10 Heating failure
track 1
Molybdenum focal track is unser-
viceable.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL
messages and contact GEMS Field Service.
You can still work with rhodium focal track.
E11 Heating failure
track 2
Rhodium focal track is unservice-
able.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
You can still work with molybdenum focal track.
E12 Checksum error
(gen)
Calibration parameters have
been modified.
Power OFF/ON from the X-ray Console. If the
problem continues, note ALL messages and con-
tact GEMS Field Service.
E13 Gen./Arm commu-
nication fail
Generator cannot communicate
with arm.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact
GEMS Field Service.