Error messages
16-errors.fm
Page no. 195
Chapter 16
X04 Collimation com-
munication failure
The Field Of View cannot be set
at the correct format:
- The Collimator is not communi-
cating.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X05 Collimation control
failure
The Field Of View cannot be set
at the correct format:
- The Collimation function is
failed.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X06 Generator commu-
nication failure
The Gantry state cannot allow an
acquisition:
- The Gantry is not communicat-
ing with the Generator.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X07 Invalid bucky type
The Gantry does not recognize
the Bucky device.
- It is not possible to perform an
image acquisition.
1) Unplug the Bucky.
2) Reinstall the Bucky: check that it is correctly
installed and switched on (initialization noise
heard).
3) Power OFF/ON from the X-ray Console.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
X09 Compression pad-
dle detection error
The Gantry does not recognize
the paddle.
- The compression thickness can-
not be computed accurately.
- It is not possible to continue the
current exam, exposure is inhib-
ited.
1) Remove and reinstall the paddle.
2) If the problem persists, power OFF/ON from the
X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
X10 X-ray beam form
software error
An error occurred during X-ray
beam format computation by the
Gantry CPU.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X11 Tube head com-
munication failure
The tube focal spot size cannot be
set:
- The tube head device is not
communicating.
- It is not possible to perform an
image acquisition.
1) If you hear the tube tilt motor, the system is try-
ing to recover itself.
2) If the message "S61 X-ray beam not ready" is
present more than 60s on the x-ray control con-
sole, power OFF/ON from the X-ray console.
3) If the problem persists, shut down the system
from the Browser.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
Message on
Gantry Readout
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).