Error messages
16-errors.fm
Page no. 167
Chapter 16
E14 Generator CPU/
INTG failure
A problem has been reported by
the Generator CPU.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E15 Console communi-
cation fail
Generator cannot communicate
with X-ray Console.
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E18 IDC/SENO Syn-
chro. fail
A problem occurred in synchroni-
zation between Image Detection
Controller system and Generator
during acquisition. The image
may be fully acquired, but dose
parameters will be missed (set to
0 in the image).
Next exposure is inhibited.
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
E19 IDC/SENO down,
DPS is OFF
Image Detection Controller sys-
tem cannot communicate with
Generator.
The Detector Power Supply has
been powered OFF by the sys-
tem.
1. Power OFF/ON from the X-ray Console.
2. If the problem occurs again at the next acquisi-
tion, power OFF the whole system and wait for the
end of the shutdown sequence.
3. Restart the system.
4. If the problem persists, note ALL messages and
inform GEMS Field Service.
E20 Check Main Distri-
bution Rack
Supply of Control Station and
Detector environment circuits is
not authorized.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
E21 Detector supply not
supplied
Supply of Detector circuits is not
authorized.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
E22 Conditioner not
supplied
Supply of detector Conditioner
circuits is not authorized.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
E23 AWS cart not sup-
plied
Supply of Control Station circuits
is not authorized.
1. Power OFF/ON from the X-ray Console.
2. If the problem persists, note ALL messages and
inform GEMS Field Service.
E24 Failure before
exposure
Current image acquisition
aborted before exposure due to
failure in the equipment
Power OFF/ON from the X-ray Console.
If the problem continues, note ALL messages and
contact GEMS Field Service.
E25 AWS/SENO Syn-
chro. fail
A synchronization problem
occurred between AWS system
and Generator during acquisition.
Next exposure is inhibited.
1. Switch OFF the whole system and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, note ALL messages and
inform GEMS Field Service.
Message on
X-ray Console
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).