Error messages
16-errors.fm
Page no. 171
Chapter 16
I32 AOP abort: use
manual mode
or
I32 Abort: Image signal
too high
This error indicates that the aver-
age signal level of the pre-expo-
sure image is too high to
calculate the X-ray parameters
for the exposure.
Switch to Manual mode, or ensure that conditions
are correct for use of AOP.
When using AOP:
- Ensure that the breast extends into the AOP
sensing zone.
- Verify that the breast thickness is shown as
greater than 0 cm on the Gantry Readout.
If the problem recurs frequently after taking the
above precautions, note ALL messages and inform
GEMS Field Service.
I33 Abort: No AOP zone
found
This error indicates that no glan-
dular region was found.
- Do not use AOP if the breast thickness exceeds
9 cm (3.5 inches).
- Verify that the breast thickness is shown as
greater than 0 cm on the Gantry Readout.
- Ensure that the breast extends into the AOP
sensing zone.
- Ensure that no highly attenuating objects (mark-
ers larger than 2 mm2, needles, etc.) are inside
that AOP sensing zone.
- Ensure that the compression force and thickness
measurements are correctly calibrated.
I34 Thickness out of
AOP range
This error indicates that AOP
Image Quality can not be
ensured, because the breast
thickness is outside the AOP
thickness range of 0 cm - 9 cm (0
to 3.5 inches).
- Do not use AOP if the breast thickness exceeds
9 cm (3.5 inches).
- Verify that the breast thickness is shown as
greater than 0 cm on the Gantry Readout.
- Ensure that the breast extends into the AOP
sensing zone
- If the breast is too thick, or it can not be posi-
tioned in the AOP sensing zone, use manual expo-
sure control.
I46 System Setting in
progress
AWS set-up in progress (loading
of Flat Field map) after a new fil-
ter was selected.
No action required.
This message is displayed for two seconds after
selection of a new filter.
Message on
X-ray Console
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).