GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm
Page no. 179
Chapter 16
AA9
Shutdown Process started
...
The AWS shutdown process
has been started because com-
munication with the UPS was
lost for more than 30 seconds.
1. Stop system usage and wait for the
end of the shutdown sequence.
2. Restart the system.
3. If the problem persists, contact
your
Field Service Representative
.
AA10 Shutdown Process started
...
The AWS shutdown process
has been started because a
UPS low battery condition was
detected (battery power must be
available to guarantee safe
operation and shutdown in the
event of power failure).
1. Stop system usage and wait for the
end of the shutdown sequence.
2. Check the mains power supply.
3. Check (on UPS panel) that the batter-
ies are charging. Do not attempt to
restart the system until the batteries
have recharged.
4. If the problem persists, contact
your
Field Service Representative
.
AA11 System is down. System
will power off within 30
seconds: do NOT reboot
now! (if system does not
turn off, switch off UPS
manually)
The message appears at the
end of the normal shutdown
sequence, when the UPS is
about to shut down, to prevent
users from trying to reboot.
No action is normally required.
If the system fails to shut down com-
pletely (the AWS
LCD
monitor continues
to display the message after one or two
minutes), switch it off by pressing the
UPS
OFF
button at the front of the Gen-
erator Cabinet.
AA12 System not ready, exam
forbidden
A system component is not yet
ready (after system start-up).
Wait until system initialization is com-
plete before requesting an acquisition.
AA13 UPS battery needs replac-
ing. Please call service
The UPS selftest has detected
that the battery requires
replacement.
Stop system usage and contact
your
Field Service Representative
.
AA14 UPS output overload.
Please call service
An UPS overload condition has
been detected.
Stop system usage and contact
your
Field Service Representative
.
BB
These messages relate to boot/reset events:
BB4
AWS reset in progress:
please wait
An AWS boot or reset sequence
is in progress.
The AWS may start an auto-
matic reset if it detects an inter-
nal acquisition device error.
Wait until system initialization is com-
plete.
If several automatic resets occur, contact
your Field Service Representative
.
BB6
Communication with UPS
not established. Please
call service. System will
power off within 30 sec-
onds: do NOT log in now!
This message appears at the
end of the power-up/boot
sequence, if communication
cannot be established with the
UPS.
1. Wait for the end of the shutdown
sequence.
2. Try to restart the system.
3. If the problem is repeated, contact
your Field Service Representative
.
BB8
Remote connection in
progress : do not acquire
images
A GE Insite connection is in
progress; no new images should
be acquired.
No action is required; wait for completion
of the connection.
Message on AWS screen
Explanation
Corrective action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).