Chapter 15
Page no. 164
15-maint.fm
GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Maintenance
3-7
Evaluation
•
For the flexible compression paddle, which is not intended to be flat and parallel to the breast support
table during compression, the deflection must be within the following values:
-
Minimum: 12 mm.
-
Maximum: 14 mm.
3-8
Corrective action
•
If the deflection exceeds the specified limits, the paddle should be replaced.
3-9
References