Chapter 16
Page no. 192
16-errors.fm
GE Healthcare
Senographe DS Acquisition System
Revision 1
Operator Manual 5307907-3-S-1EN
Error messages
W06 Invalid language.
Default is used
The Gantry CPU received incor-
rect language information and will
use the default value.
The default language value is the
English language.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, continue to use the sys-
tem with English language set. Note ALL mes-
sages and contact GEMS Field Service.
W07 Control buttons
error
An error occurred on an arm con-
trol button keypad, which does not
affect the current system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
W08 A footswitch is
pressed
An error occurred on a footswitch
device that does not affect the
current system use.
Note ALL messages and inform GEMS Field Ser-
vice at next field service maintenance.
W09 Gantry Readout
failure
The gantry readout has failed.
Contact GEMS Field Service.
W10 Lift calibration
error
The lift device is not well cali-
brated; no movement will be avail-
able until the lift device will be well
calibrated.
Note ALL messages and contact GEMS Field Ser-
vice.
W11 Lift control error
An error occurred during lift move-
ment that does not affect the
safety control but the speed pro-
file only.
Note ALL messages and contact GEMS Field Ser-
vice.
W12 Lift end position
exeeded
The Lift movement has stopped at
a firmware limit, but should have
stopped before.
1) Use lift slow speed in the upper and lower limit
area.
2) Note ALL messages and contact GEMS Field
Service.
W15 Rotation calibra-
tion error
The rotation device is not cor-
rectly calibrated; no movement
will be available until the rotation
device has been correctly cali-
brated.
Note ALL messages and contact GEMS Field Ser-
vice.
W16 Angulation control
error
An error occurred on the Rotation/
Angulation clutch/brake control.
1) If an exam is currently in progress continue it,
using the same rotation incidence.
2) Move away from 0° and return to 0°; try again to
switch to rotation mode.
3) If the problem persists, move the compression
paddle up, switch to rotation mode, perform rota-
tion, and return to angulation mode.
4) If the problem persists, power OFF/ON from the
X-ray Console.
5) If the problem persists, note ALL messages and
contact GEMS Field Service.
W17 Rotation control
error
An error occurred during rotation
movement. It does not affect the
safety control, but only the speed
profile.
Note ALL messages and contact GEMS Field Ser-
vice.
Message on
Gantry Readout
Explanation
Corrective Action
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).