Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-4
Software Loading Work-flow and Check-list
This check sheet is prepared to assist a FE during software loading.
Customer Name
Customer Contact
Console #
Unit ID
Unit Type
Section
Item
Check Parameter
Result
5-2-2-2
BT09_SV_BEP4
Configuration Varia-
tionos
Check current system version before ghost/application
loading
Should be R8.x.x or later
Ver.
8-2-6-1 - Disabling
USB Devices
Check if all USB printers are OFF.
Check if SCSI UP-D50 is ON.
Should be OFF.
OK
N/A
8-2-6-2 - Save Cus-
tomer Data / Setting
Save Customer Setting
User and Password
OK
N/A
Save Customer Setting
Connectivity Parameter
OK
N/A
Save Customer Setting
Option Key String
OK
N/A
Regional Preset
OK
N/A
Install Base System
Select fullghost or C ghost
Full
C only
8-2-8 - Modify Sys-
tem/Windows Set-
tings
UPS Setting
Run UPSSETUP.reg
OK
NG
Drive Letter setting
CD/DVD to be Drive ‘G’
OK
NG
Hardware Divice
OK
NG
LAN Parameter
OK
NG
TCP/IP Setting
Verify
OK
NG
Date/Time Setting
Verify
OK
NG
8-2-9 - Installing
R8.x.x Application
Software
Application CD correctly recognized
App loading
OK
NG
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......