Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-8-2
Drive Letter Setting
If you CANNOT see this dialog at the first boot up, this dialog appears at 2nd boot up.
Please choose
OK.
You should check the Drive Letter manually by Windows Explorer for double check.
-
DVD Drive should be assigned to G:
If Drive letter is wrong, please double-click
Remap Drive Letter
icon on the desktop.
8-2-8-3
Hardware Device Check:
1.) On the windows desktop, left-click on
My Computer
, then right-click on
Properties.
Figure 8-70 OK
CAUTION
Drive F: (Swap) Letter is removed on R7.0.0EZ. (This is correct.)
Figure 8-71 Properties