Section 5-5 - Common Service Platform
Temperature
Error (!!)
The error occurred in the temperature with “!!” mark. In the time when the error occurred, the
message “Temp is above safe limit. Check air-filter! Shutdown in #seconds.” appears in the status
bar and the system is automatically shut down.
The temperature which the error occurs:
•
75 degrees or higher for Rack Temperature (1)
•
65 degrees or higher for Rack Temperature (2)
•
65 degrees or higher for HV Temperature
Warning (@@)
The warning occurred in the temperature with “@@” mark. In the time when the warning occurred,
the message “This system is overheated! Check air-filter!” appears in the status bar.
The temperature which the warning occurs:
•
70 degrees or higher for Rack Temperature (1)
•
60 degrees or higher for Rack Temperature (2)
•
60 degrees or higher for HV Temperature
Info (^^)
The information occurred in the temperature with “^^” mark to call attention. In the time when the
information occurred, the message “Approaching the overheat limit! Check air-filter!” appears in the
status bar.
The temperature which the information occurs:
•
65 degrees or higher for Rack Temperature (1)
•
50 degrees or higher for Rack Temperature (2)
•
50 degrees or higher for HV Temperature
Others
Temperature information
Probe
Displays the probe name and connector #.
Board
Displays the Part #, Board #, and Dip switch revision.
DICOM
Whenever DICOM data are sent, logs are reported. An error log is reported in an error status and an information log is
reported in a normal status. (The log is command or status information specified by DICOM statement.)
Logs
Descriptions
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......