Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-25
7.) If your system has no TCP port numbers as follows, perform these procedure to add the TCP port
numbers. If the TCP port already have been set, skip these procedures. (This depends on the
network card installed in the PC box.)
*
At first check mark
Enable TCP/IP Filtering (All adapters)
.
*
Check mark
Permit Only
in TCP ports box
*
Click
Add...
button.
*
Enter
21
(port number) as the TCP Port ,then click
OK
button to apply it.
*
Repeat these procedures to register
21, 23, 80, 104, 512, 3003, 5800, 5801, 5900, and
5901
for TCP ports.
Figure 8-81 TCP port number
NOTICE
UDP Ports and IP Protocols must be kept default settings.
The TCP port numbers,
21, 23, 80, 104, 512, 3003,
5800, 5801, 5900, and 5901
, have been set.
No TCP port number is set. They must
be set.
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......