Section 4-2 - General Procedure
4-2-2-3-2
R 4.x.x to R6.0.x
1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
2.) The “Probing for maintenance access” dialog appears. Enter
Password
, then click on
OK.
3.) Then Start Loader display will be shown on the Monitor display.Click on
Maintenance.
4.) Click on
Exit to Windows
in maintenace diaglog window.Then the scanner is booted up
automatically.
4-2-2-3-3
R2.x.x to R4.x.x
1.) Insert the service dongle in the Service port located at the rear panel.
2.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
3.) Then Start Loader display will be shown on the Monitor display. Then the scanner is booted up
automatically.
NOTE:
To enter the Maintenance Mode, select Maintenance button.
NOTE:
Start is selected automatically when it time out.
Figure 4-8 Clicking on Mainttenance
Figure 4-9 Start Application Window
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......