GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
6-14
Section 6-5 - Monitor and LCD Adjustments
6-5-4
LCD Monitor Resolution
Should one encounter “small screen” on the LCD monitor, screen resolution must be adjusted in order
to view full size.
1.) From the touch panel, select
Utility > System > About.
2.) Click
Additional About Info.
3.) Select
High Resolution.
6-5-5
LCD Touch Panel Adjustment
There are two types of the touch panel LCD in LOGIQ 7 (manufactured by Gaudi or Technart).
Depending on the LCD type on the system, adjustment procedure is slightly different. Some models of
LOGIQ 7 have the LCD with the cover plate. Open the key board and remove those plate to identify the
content:
This adjustment must be performed when always replacing:
-
PC Box Assy
-
Keyboard Assy
-
Cable between LCD unit and PC Box
Figure 6-15 LOGO Location
Gaudi Logo
Technart Logo
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......