Section 6-5 - Monitor and LCD Adjustments
6-15
6-5-5-1
For Gaudi LCD
1.) Refer to
to enter the windows desktop.
2.) The following screen should appear when exiting to Windows.
3.) Check the followings:
-
Proper balance of the LCD display as shown in Figure 6-17
-
No Jitter (Phenomenon that a character does not move sideways)
NOTE:
If the location of the LCD screen is appropriately set in center, do not proceed this adjustment
procedures.
4.) Open the OP Panel, and hold the OP Panel using maintenance service rod. Refer to Replacement
Procedure in chapter 8 for the procedures.
Figure 6-16 Start-up Screen
Figure 6-17 Example of LCD Bad balance
NOTICE
Only when the windows desktop appears in the 17 inch CRT monitor or 17 inch LCD monitor, you can
use Auto Adjust function. Also 640 x 480 dots pictures must be displayed on the LCD monitor to function
Auto Adjust correctly.
Dislocation
Example
LCD Display
Display
Screen
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......