GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-29
8-2-9
Installing R8.x.x Application Software
8-2-9-1
Installing R8.x.x Application Software
1.) Power ON the scanner. The windows desktop appears.
2.) Insert the Application Software CD-ROM into the DVD drive.
NOTICE
You must install BT09 Base image before Application Software installation. If not, Application shows the
error message, “Service platform is out of date. You must reghost the C drive and reinstall the software.”
NOTICE
If you are attempting to install only application software, you MUST first check current Regional Preset
Value. Press
Utility > System > About
(Details in 4-2-6 - Archiving and Loading Presets for BT07
Figure 8-87 Desktop
NOTICE
At this time, the DVD writing software screen appears, but you can ignore the screen.
NOTICE
Wait approx. 20 seconds until the disk can be read.
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......