Section 2-2 - General Console Requirements
2-5
2-2-3
EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ™ 7 complies with
limits as stated on the EMC label. However there is no guarantee that interface will not occur in a
particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of defect.
These sources include:
•
medical lasers,
•
scanners,
•
cauterizing guns,
•
computers,
•
monitors,
•
fans,
•
gel warmers,
•
microwave ovens,
•
light dimmers,
•
portable phones.
The presence of broadcast station or broadcast van may also cause interference. See for EMI
Prevention tips.
Table 2-5 EMI Prevention/abatement
EMI Rule
Details
Be aware of RF sources
Keep the unit at least 5 meters or 15 feet away from other EMI sources. Special shielding
may be required to eliminate interference problems caused by high frequency, high
powered radio or video broadcast signals.
Ground the unit
Poor grounding is the most likely reason a unit will have noisy images. Check grounding
of the power cord and power outlet.
Replace all screws, RF
gaskets, covers, cores
After you finish repairing or updating the system, replace all covers and tighten all
screws. Any cable with an external connection requires a magnet wrap at each end.
Install the shield over the front of card cage. Loose or missing covers or RF gaskets allow
radio frequencies to interface with the ultrasound signals.
Replace broken RF gaskets
If more than 20% or a pair of fingers on the RF gaskets are broken, replace the gaskets.
Do not turn on the unit until any loose metallic part is removed.
Do not place labels where
RF gaskets touch metal
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will
permit RF leakage. Or, if a label has been found in such a position, move the label.
Use GE specified
harnesses and peripherals
The interconnect cables are grounded and require ferrite beads and other shielding.
Also, cable length, material, and routing are all important; do not change from what is
specified.
Take care with cellular
phones
Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Properly dress peripheral
cables
Do not allow cables to lie across the top of the card cage or hang out of the peripheral
bays. Loop the excess length for peripheral cables inside the peripheral bays. Attach the
monitor cables to the frame.
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......