Section 6-5 - Monitor and LCD Adjustments
5.) Find out the notch to access OSD Switch located at backside of the LCD as shown in Figure 6-18.
Press the switch inside of the notch.
NOTE:
Be careful not to cut fingers with notch.
6.) OSD menu is shown on the LCD display.
7.) Remove the maintenance rod, and re-install the OP Panel temporarily to use the LCD touch panel.
Refer to Replacement Procedure in chapter 8 for the procedures.
8.) Using
SELECT
button, select
AUTO ADJUST
. The letters will be blue when it is selected.
9.) Make sure that the
AUTO ADJUST
is selected, then touch
ENTER
.
10.)LCD display Auto Adjustment procedures will be started automatically.
Figure 6-18 Switch Location
NOTICE
Never select AUTO CONTRAST in this menu. Selecting AUTO CONTRAST might occur in an improper
contrast.
Figure 6-19 OSD Softmenu 1
Notch
to access
OSD SW
10A5V1.0
MANUAL ADJUST
AUTO ADJUST
AUTO CONTRAST
EXIT
SELECT ENTER
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......