Section 5-3 - Peripheral Compatibility
5-11
Section 5-3
Peripheral Compatibility
NOTE:
For SONY D21MD (or D23MD) color printer ONLY:
The D21MD (or D23MD) printer contains the L-sized ribbon cassette and paper as a standard.
However, the S-sized paper is recommended for LOGIQ 7 system. If the S-sized paper is used
with the L-sized ribbon cassette, “Incorrect paper size setting” error occurs with no print.
So, when using the S-sized paper, change the ribbon cassette to the S-sized.
Figure 5-35 Peripheral Compatibility