LOGIQ™ 7 S
ERVICE
M
ANUAL
8-10
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-2-6-4
Saving Connectivity
Before starting upgrading, write down the
Connectivity Setting
for back-up.
1.) Touch
Utility > Connectivity
on the touch panel and click the
TCP/IP
tab.
2.) Write down the following parameters:
Figure 8-60 TCP/IP
Parameters
Descriptions (Entry for Stand-Alone System)
Computer Name
Enable DHCP
This parameter
must NOT
be selected.
IP-Address
(192.168.1.2)
Subnet Mask
(255.255.255.0)
Default Gateway
(192.168.1.1)
Note : For a stand-alone system, preset values of IP adress, subnet mask, and default gateway, shown above,
must entered.
Table 8-3 TCP/IP
Remove the
Check mark.