GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
2-6
Section 2-2 - General Console Requirements
2-2-4
Probes Environmental Requirements
Table 2-6 Operation and storage Temperatures for 2D Probes
Operational
Storage
Transport
Temperature
10 - 40
o
C
50 - 104
o
F
-10 - 60
o
C
14 - 140
o
F
-40 - 60
o
C
-40 - 140
o
F
Humidity
30 - 85%
non-condensing
30 - 90%
non-condensing
30 - 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Table 2-7 Operation and storage Temperatures for 4D Probes
Operational
Storage
Transport
Temperature
18 - 30
o
C
64- 86
o
F
-10 - 50
o
C
14 - 122
o
F
-10 - 50
o
C
14 - 122
o
F
Humidity
Max. 70%
non-condensing
Max. 90%
non-condensing
Max. 90%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......