Section 4-6 - Power Supply
4-29
Section 4-6
Power Supply
4-6-1
Power Supply Test Procedure
There is no need to perform any special tests on the Power Supplies if there does not seem to be a
problem that may be related to the Power Supplies.
Refer to Section 6-2 - LV Unit Adjustments (For BT03 or lower) if you appear to have a problem that
may be related to the Power Supplies.
4-6-2
Power Supply Adjustment
There are no adjustments on the power supplies. The DC Power is self-regulated. If a voltage is outside
the specified range, it means that something is wrong, either with the power supply itself or with a unit
connected to that specific power outlet.