Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-51
8-2-10-10
Regional Preset
9.) Touch
Utility > System
.
10.)Click
About
tab.
11.)Click
Additional About Information
.
12.)Check the region name of
Preset Region (Rquires reboot):
If it is None, skip this section.
13.)Click
Reload Factory Default
button.
14.)Click
OK
on confirmation dialog box.
This changes Preset Region (Requires reboot) to
None
.
NOTICE
For the system other than "NONE" selected as Regional Preset, resetting MUST be required.
Figure 8-113 Preset Region
Figure 8-114 Reload Factory Default