Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-45
Regional Setup (For the system with language other than English ONLY)
4.) Select the
Advanced tab
, then select
Russian or Greek
in the Language for non-Unicode
programs pull-down menu. Press
Apply
. Answer
Yes
to use files already loaded on the hard disk,
then answer
No
to not reboot the system yet, press
OK
. Press
Save
and Exit the Utility screen.
5.) Reboot the system. When your system restarts, the system appears in the selected language.
6.) To switch between the English and Russian (or Greek) keyboard, press
Alt
+
Shift
to change the
keyboard to the Russian or Greek keyboard.
7.) Apply the changes by pressing
Apply
. Press
OK
TWICE
.
Figure 8-107 Advanced tab
NOTICE
To have the settings take effect, you MUST turn off the system and turn it back on.
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......