GE H
EALTHCARE
D
IRECTION
2286865, R
EVISION
14
LOGIQ™ 7 S
ERVICE
M
ANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
8-17
8-2-8
Modify System/Windows Settings
After Base System is successfully installed, various parameters must be properly set before proceeding to Application
load. The below outlines necessary system/windows setting.
8-2-8-1
UPS Setting
UPS automatically setup by Ghost. You do not need to setup it manually. You can see the message
below on the desktop.
Please choose
OK
.
Table 8-6 System/Windows Settings
Item
Method
Proper Value
UPS Setting
Automatic Setting
(n/a – just load)
(Restart System)
Avance AC97 Audio
Load from c:\windows\system32\drivers
(n/a – just load)
(Restart System)
Drive Letter
Automatic Setting but check desktop icon if not.
DVD Drive = G
HW Device
My Computer
\ Properties
Various
LAN
Desktop
\ My Network Places
\ Properties
\ Local Area Connection
Double Click to activate
TCP/IP Setting
Desktop
\ My Network Places
\ Properties
\ Local Area Connection
\ Internet Protocol (TCP/IP)
\ Properties
\ Advanced
\ Options
\ TCP/IP Filter
Values in ‘Permit Only’
Date/Time
Start
\ Settings
\ Control Panel
\ Date and Time
Proper Time Zone
Accurate Date / Time
Figure 8-69 UPS Setting
Содержание LOGIQ 7
Страница 2: ......
Страница 9: ... vii GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL ...
Страница 12: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL x ...
Страница 84: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 3 26 Section 3 5 Installation Paperwork ...
Страница 116: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 4 32 Section 4 7 Site Log ...
Страница 204: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 6 42 Section 6 8 Jumper and Dip Switch Setting ...
Страница 392: ...GE HEALTHCARE DIRECTION 2286865 REVISION 14 LOGIQ 7 SERVICE MANUAL 10 34 Section 10 7 Electrical Safety Tests ...
Страница 393: ......