Section 6-5 - Monitor and LCD Adjustments
6-21
6-5-5-2-2
Manual Deblurring (For Technart ONLY)
Perform manual deblurring only when auto adjustment fails.
1.) Refer to
to enter the windows desktop.
2.) Select
Program > Accessories > Paint
to run the paint software.
3.) Phase Check:
a.) Drag and drop the paint screen on the touch panel screen.
b.) Click on square symbol at the top right corner of the paint screen to maximize the screen.
Figure 6-26 Paint Software
Figure 6-27 Moving Paint Screen